132
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
“Maintenance” Sub-menu Functions
11.14.2
Changing the time zone
Function
With this menu the internal clock of the EuroFID can be adjusted by the hour. This is an
alternative method to set the internal time to your local time.
The standard setting is “+1.00” = +1 hour later than Greenwich Mean Time (GMT) =
Central European time (CET).
Procedure
1
Open the menu: Main menu
Maintenance
System Parameters
User level
Next
menu
Change time zone
2
Press
or .
➝
The entry field is highlighted.
3
Enter the required value (possible:
- 1 2 . 0 0
to
+ 1 2 . 0 0
).
4
Press
or
.
➝
E n d
is highlighted.
5
Press .
➝
The internal clock then starts with the set values.
Switching from summer/winter time
S y s t e m P a r a m e t e r s
C h a n g e t i m e z o n e
T i m e z o n e
+ 1 . 0 0 H
C h a n g e o r
6 1 . 6 m g C
R a n g e . . . 1 0 0
Menu title
Entry field
Current measured value
Current measuring range (measuring range end value)
To enter a negative value:
1
Press the plus/minus key [+/
–
].
➝
The entry field switches to
+ 0 . 0 0
.
2
Enter the required value.
➝
The value appears with a negative sign.
E n d
Содержание EuroFID
Страница 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Страница 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Страница 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Страница 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Страница 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Страница 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Страница 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Страница 155: ...Shutdown EuroFID Operating Instructions 8011094 V2 3 SICK AG 155 EuroFID 13 Shutdown Preparations Shutdown procedure ...
Страница 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Страница 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...