“Maintenance” Sub-menu Functions
11.11.3
Setup filter (averaging for the “Measured value” analog output)
Function
It is possible that a plotting device is connected to the “Measured value” analog output
which does not record the measured value continuously, but only at regular intervals – e.g.
a printing recorder. It can be advantageous to average the analog measurement signal over
the respective time interval; otherwise the plotter would register momentary values which
may not be representative.
To achieve this effect, the “Setup filter” can be programmed.The setting value is the time
constant of sliding averaging (time-related integration).
Procedure
1
Open the menu: Main menu
Maintenance
System Parameters
User level
I/O-
Parameters
Setup filter
2
Press
or
.
➝
Entry field is highlighted.
3
Enter the required value.
4
Press
or
.
➝
E n d
is highlighted.
5
Press .
●
The “Setup filter” setting only affects the “Measured value” analog output
(
).
●
Standard value for emission measurement: 30 s.
There is also a similar setting option
– for the measurement display
– for the “Reference value” analog output
S y s t e m P a r a m e t e r s
S e t u p f i l t e r
I n t e g r a t i o n t i m e
3 0 . 0 s
u p : 0 . 0 s
u p t o : 6 0 0 . 0 s
C h a n g e o r
6 1 . 6 m g C
R a n g e . . . 1 0 0
Menu title
Menu title
Entry field
Minimum value
Maximum value
Current measured value
Current measuring range (measuring range end value)
E n d
Содержание EuroFID
Страница 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Страница 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Страница 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Страница 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Страница 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Страница 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Страница 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Страница 155: ...Shutdown EuroFID Operating Instructions 8011094 V2 3 SICK AG 155 EuroFID 13 Shutdown Preparations Shutdown procedure ...
Страница 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Страница 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...