Clearing Malfunctions
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
173
15.4
Overtemperature protection on the analyzer unit
15.4.1
Function of the overtemperature protection
To protect the measuring system against damage caused by overheating, the analyzer unit
has a thermoelectric temperature monitor (thermostat) which acts as an overtemperature
protection device. If the temperature in the analyzer unit becomes too high, the
temperature monitor automatically switches off the measuring system and heating of the
analyzer unit.
15.4.2
Resetting the temperature monitor (after tripping)
If the temperature monitor has been tripped:
1
Leave the analyzer unit to cool.
2
Open the housing of the analyzer unit.
3
Press the reset switch on the temperature monitor (
).
4
Close the analyzer unit.
Fig. 31
Temperature monitor in the analyzer unit
NOTICE:
When the operating temperature is exceeded,
– the measured values can be corrupted
– Components of the measuring system can be damaged
– The analyzer unit could become dangerous to use.
Do not change the setting of the temperature monitor.
Have the function of the temperature monitor checked at least once a year
(
).
NOTICE:
On devices which are used as a gas detector, the overtemperature protection
ensures compliance with the classification temperature (Refer to the type plate
for the specification).
Do not change the setting of the temperature monitor.
Reset switch
50 °C
300 °C
Содержание EuroFID
Страница 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Страница 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Страница 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Страница 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Страница 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Страница 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Страница 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Страница 155: ...Shutdown EuroFID Operating Instructions 8011094 V2 3 SICK AG 155 EuroFID 13 Shutdown Preparations Shutdown procedure ...
Страница 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Страница 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...