In the setting “toggle between on and standby,” the balance power will turn back
on automatically after a power failure has occurred or after you have disconnected
your balance temporarily from AC power.
In the setting “automatic power-on,” the balance will turn back on automatically
after a power failure has occurred; after the balance has been disconnected, then
reconnected to AC power; or after the
e
key has been pressed. In this setting,
the balance can no longer be turned off using the
e
key.
Power-on mode
Code
(Power) off –> on <–> standby
* 8 6 1
On <–> (power) off
8 6 2
On <–> standby
8 6 3
Automatic power-on
8 6 4
Automatic Shutoff (Battery Saver) for Use with Code 8 6 2 or the
Rechargeable Battery Option**
If you are using the rechargeable battery option, turn the automatic shutoff function
on to increase the hours of battery operation so that you do not need to recharge
the batteries that often. Here is how the battery saver function works: If there has
not been a change in a weight readout for at least 2 minutes or you have not
pressed any keys during this time, “
b
” will be displayed. After another 2 minutes,
the balance will shut off if it has not been used in the meantime.
Automatic shutoff
Code
On
8 7 1
Off
* 8 7 2
Display Backlighting
Depending on your individual workplace requirements, you can turn the display
backlighting on or off accordingly.
Display backlighting
Code
On
* 8 8 1
Off
8 8 2
*
*
= factory setting
** = not applicable to verified balances approved for use as legal
measuring instruments
2–20
Содержание AC 121S-00MS
Страница 91: ...Sartorius Master Series Product Data Sheets 6 1...
Страница 105: ...Dimensions Scale Drawings AC 211S AC 211P AC 121S 0CE Note All dimensions are given in millimeters 6 24...
Страница 107: ...LC 2201S LC 2201P LC 821 0CE LC 421 GC 1201 G Note All dimensions are given in millimeters 6 26...
Страница 108: ...LC 16000 S LC 34000 P LC 34 0CE Note All dimensions are given in millimeters 6 27...