When data are output without decimals, the decimal point is suppressed
(except when a certain display mode is selected).
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16
Data output example: +1255.7 g
1
2
3
4
5
6
7
8
9 10 11 12 13 14 15 16
Characters:
1st
Plus or minus sign or space
2nd
Space
3rd –10th Weight with a decimal point; leading zeros = space
11th
Space
12th –14th Unit symbol or space
15th
Carriage return (CR)
16th
Line feed (LF)
If the weighing system has not stabilized (stability symbol is not displayed),
no unit symbol will be output.
For the display format parameter “last numeral off or only at stability”
(blank last numeral by a rounding factor of 10 or display last numeral when the
readout is stable), the 10th character in the string is a space. The reason:
the last numeral without a decimal point is not blanked. Instead, it is given a
definite zero; otherwise, the display would go completely blank when the balance
is unloaded. In this case, you will obtain the standard display format.
* = space, U = unit
+
*
*
*
1
2
5
5
.
7
*
g
*
* CR
LF
+
*
*
*
*
*
*
– – – – – – – – – – – – – – – – – – – – –
*
*
*
*
*
* 10
6
10
5
10
4
10
3
10
2
10
1
10
0
*
CR
LF
– – – – – – – – – – – – – – – – – – – – –
U
U
U
–
0
0
0
0
0
0
5–3
Содержание AC 121S-00MS
Страница 91: ...Sartorius Master Series Product Data Sheets 6 1...
Страница 105: ...Dimensions Scale Drawings AC 211S AC 211P AC 121S 0CE Note All dimensions are given in millimeters 6 24...
Страница 107: ...LC 2201S LC 2201P LC 821 0CE LC 421 GC 1201 G Note All dimensions are given in millimeters 6 26...
Страница 108: ...LC 16000 S LC 34000 P LC 34 0CE Note All dimensions are given in millimeters 6 27...