Counting
...
...
A weight readout is stored as a reference sample quantity (factory setting: 10 pcs).
To change this value, see the description on page 4–14. (The reference sample
quantity is displayed as long as the reference sample weight has not yet been
stored for counting.) While counting (piece count readout), you can press
F
to
update the reference sample for optimizing the accuracy (approx. double the
reference sample quantity). In the process, the reference weight is recalculated
based on the total reference sample weight and the piece count displayed.
Counting Small Parts
Code 2 1 4
Menu codes used in the example:
Function
Code
Counting
2 1 4
With data ID code
7 2 2
Ref. qu ref. weight
7 1 2
Application: counting of bulk quantities of items with the same weight
Step/(key ID)/key
Readout
Data output/printout
Place container on the pan
+
22.65 g
c
,
t
0.00 g
Add 10 sample parts
+
58.55 g
Store: (STO REF)
F
+
10 pcs
nRef
+
10 pcs
wRef
+
5.8550 g
To update ref. sample,
add more parts
(approx. 2x nRef)
+
20 pcs
press (OPT)
F
+
5.8544 g
wRef
+
5.8544 g
Fill container with
desired quantity
+
286 pcs
p
+
286 pcs
Qnt
+
286 pcs
To obtain weight
readout: press
w
+ 1676.66 g
p
+ 1676.66 g
N
+
1676.66 g
The data ID codes
mean:
nRef
+
10 pcs
reference sample quantity (piece count)
wRef
+ 5.8550 g
reference weight
Qnt
+
286 pcs
calculated piece count
N
+1676.66 g
net weight
As an alternative, you can
select this output parameter:
Reference weight
7 1 3
4–13
Содержание AC 121S-00MS
Страница 91: ...Sartorius Master Series Product Data Sheets 6 1...
Страница 105: ...Dimensions Scale Drawings AC 211S AC 211P AC 121S 0CE Note All dimensions are given in millimeters 6 24...
Страница 107: ...LC 2201S LC 2201P LC 821 0CE LC 421 GC 1201 G Note All dimensions are given in millimeters 6 26...
Страница 108: ...LC 16000 S LC 34000 P LC 34 0CE Note All dimensions are given in millimeters 6 27...