for rectangular weighing pans with a capacity <12 kg
Remove the following parts from the balance:
– Weighing pan
(1)
– Pan draft shield
(28)
(depending on the model)
Detach the dust cover by peeling it off at the adhesive
strips. Remove all traces of adhesive.
Then remove the white backing from the adhesive
strips on the new dust cover. Attach the dust
cover by
gently pressing down on the adhesive strips.
Safety Inspection
If there is any indication that safe operation of the
balance with the AC adapter is no longer warranted,
turn off the power and disconnect the equipment from
AC power immediately. Lock the equipment in a secure
place to ensure that it cannot be used for the time being.
Safe operation of the balance with the AC adapter
is no longer ensured when
– there is visible damage to the AC adapter
– the AC adapter no longer functions properly
– the AC adapter has been stored for relatively
long periods under unfavorable conditions.
In this case, notify your nearest Sartorius Service
Center or the International Technical Support
Department based in Goettingen, Germany. Only
service technicians who are authorized by Sartorius
and have access to the required maintenance manuals
are allowed to perform maintenance and repairwork
on the equipment.
We recommend that the AC adapter be regularly
inspected by a qualified Sartorius service technician
according to the following checklist:
– Insulation resistance >7 megohms measured
with a constant voltage of at least 500 V
at a 500 kohm load
– Equivalent leakage current <0.05 mA measured by
a properly calibrated multimeter
1–31
Содержание AC 121S-00MS
Страница 91: ...Sartorius Master Series Product Data Sheets 6 1...
Страница 105: ...Dimensions Scale Drawings AC 211S AC 211P AC 121S 0CE Note All dimensions are given in millimeters 6 24...
Страница 107: ...LC 2201S LC 2201P LC 821 0CE LC 421 GC 1201 G Note All dimensions are given in millimeters 6 26...
Страница 108: ...LC 16000 S LC 34000 P LC 34 0CE Note All dimensions are given in millimeters 6 27...