Net Total
Code 2 2 2
Press
f
(identified by “TAR” in the display) to store each component that you
weigh-in. The balance is tared automatically each time you store a weight.
Therefore, you can weigh-in the components one after the other starting with
a zero readout.
Menu codes used in the example:
Function
Code
Tare memory
2 2 2
With data ID code
7 2 2
Automatic data output
7 1 2
Printout of individual
components/tare weights
*7 3 1
Application: simple compounding and formulating in the lab or in production
Step/(key ID)/key
Readout
Data output/printout
Place container on
the pan
+
22.65 g
c
,
t
0.00 g
Weigh-in 1st component
+
4.61 g
Store: (TAR)
f
0.00 g Net
N1
+
4.61 g
Weigh-in 2nd component
+
60.33 g
Store: (TAR)
f
0.00 g Net
N1
+
60.33 g
Weigh-in and store
additional components
+
xxx.xx g Net
N1
+
xxx.xx g
Finish weighing-in
procedure and output
the total weight
c
,
p
+
272.89 g
N
+
272.89 g
The data ID codes
mean:
N1
+
60.33 g
net weight of stored component
N
+
272.89 g
net total weight
*= factory setting
As an alternative, you can select
this output parameter:
Printout of net total weight/
total tare weight
7 3 2
4–5
Содержание AC 121S-00MS
Страница 91: ...Sartorius Master Series Product Data Sheets 6 1...
Страница 105: ...Dimensions Scale Drawings AC 211S AC 211P AC 121S 0CE Note All dimensions are given in millimeters 6 24...
Страница 107: ...LC 2201S LC 2201P LC 821 0CE LC 421 GC 1201 G Note All dimensions are given in millimeters 6 26...
Страница 108: ...LC 16000 S LC 34000 P LC 34 0CE Note All dimensions are given in millimeters 6 27...