Over/Under Checkweighing with Additional Requirements
Changing the Tolerance Limits
You can change the over/under tolerances between 0.1% and10%.
Tolerance limits
Code
0.1% over/under deviation from the target
4 1 1
0.2% over/under deviation from the target
4 1 2
0.5% over/under deviation from the target
4 1 3
1.0% over/under deviation from the target
* 4 1 4
1.5% over/under deviation from the target
4 1 5
2.0% over/under deviation from the target
4 1 6
2.5% over/under deviation from the target
4 1 7
3.0% over/under deviation from the target
4 1 8
5.0% over/under deviation from the target
4 1 9
10.0% over/under deviation from the target
4 1 10
Analog Display
You can use an analog marker (two segments) instead of the bar graph located
below the checkweighing indication.
Analog display
Code
Off
8 5 1
Bar graph
* 8 5 2
Analog marker
8 5 3
Automatic Data Output
“Acceptable” values can be printed out or transferred automatically. Data are
output once the readout has stabilized within the tolerance limits selected.
Following data output, this function will be blocked until the balance has been
completely unloaded.
Automatic data output
Code
On
4 2 1
Off
* 4 2 2
* = factory setting
4–12
Содержание AC 121S-00MS
Страница 91: ...Sartorius Master Series Product Data Sheets 6 1...
Страница 105: ...Dimensions Scale Drawings AC 211S AC 211P AC 121S 0CE Note All dimensions are given in millimeters 6 24...
Страница 107: ...LC 2201S LC 2201P LC 821 0CE LC 421 GC 1201 G Note All dimensions are given in millimeters 6 26...
Страница 108: ...LC 16000 S LC 34000 P LC 34 0CE Note All dimensions are given in millimeters 6 27...