Interfacing Devices with the Balance
Please note that the interface port is electrically
connected to the protective grounding conductor
of the balance housing. The interface cables supplied
as standard equipment are shielded, and both ends
of each cable are electrically connected to the
connector cases. This connection may result in inter-
ference caused by ground loops or by transient
currents if you have grounded the housing or
connected the protective grounding conductor
for line power. If necessary, connect an equipotential
bonding conductor to the balance.
1–26