background image

 

 
 
 
 

Range

r™ 2000 Series 

Instruction Manual 

 
 
 
 
 
 

 

 
 
 

Summary of Contents for Ranger R21PE15

Page 1: ...Ranger 2000 Series Instruction Manual ...

Page 2: ......

Page 3: ...roved accessories and peripherals Operate the scale only under ambient conditions specified in these instructions Disconnect the scale from the power supply when cleaning Do not operate the scale in hazardous or unstable environments Do not immerse the scale in water or other liquids Do not place the scale upside down on the platform Only use weights within the scale s capacity as specified in the...

Page 4: ... in a small round window on the front of the scale To level the scale adjust the feet so the bubble is centered in the circle Be sure the equipment is level each time its location is changed Figure 2 2 Level indicator 2 5 Connecting Power AC power is used to power the scale when battery power is not needed First connect the AC power cord supplied to the power input jack then connect the AC plug to...

Page 5: ...y symbol see item 9 in table 3 2 indicates the battery charging status When charging the symbol will blink slowly and when fully charged the symbol will be turned off TABLE 2 2 Symbol Charge level Battery in use Symbol displayed Notes When battery symbol blinks fast approximately 30 minutes working time is left When lo bat is blinking the scale will shut off Charging the scale must be performed in...

Page 6: ...Toggles through active weighing units Short Press in Menu Toggles through available settings Function Mode Back Short Press Selects function setting Long Press Selects active mode Short Press in Menu Returns to previous setting M Menu Exit Short Press Accumulates the weight or displays the accumulated information at 0 load Long Press Enters User Menu Short Press in Menu Quickly exit User Menu Tare...

Page 7: ...bol 2 Center of Zero symbol 9 Battery charge symbol 3 Negative symbol 10 Pound Ounce Pound Ounce symbols 4 Stable weight symbol 11 Kilogram gram symbols 5 Preset Tare Tare symbols 12 Pieces symbol 6 Pointer symbols 13 Percent symbol 7 Accumulation symbol Figure 3 2 Display 1 2 3 4 5 6 7 8 9 10 11 12 13 ...

Page 8: ...t Yellow Loads Lower limit Figure 3 4 Below view of Ranger 2000 3 2 Turning Scale On Off To turn the scale on press and hold the On Zero Off button for 1 second The scale performs a display test momentarily displays the software version and then enters the active weighing mode To turn the scale off press and hold the On Zero Off button until OFF is displayed Lock Switch Power Input Weigh Below Acc...

Page 9: ...erence weight The display shows 100 6 Replace the reference material with the sample material The display shows the percentage of the sample compared to reference weight 7 To clear the stored reference press and hold Mode until Percnt is displayed Press Yes when Clr ref is displayed Note Press Function to view the current reference weight 3 5 Counting Mode This mode counts large numbers of items b...

Page 10: ...t and highlight the next digit Repeat until all the digits are correct Press Yes to accept the low limit value Set Hi will be displayed 4 Repeat the same procedure to accept or edit the high value 5 If required place an empty container on the pan and press Tare 6 Place sample material on the pan or in the container If the sample weight is under the target weight range the yellow LED will light If ...

Page 11: ...ings please see section 4 2 for details 3 6 3 Check Count Set Checkweigh to Count in the Mode menu Use this mode to compare the quantity of items to a target quantity range 1 Press and hold Mode until CHeCk Check is displayed CLr PwW clear APW will then display If no APW exists the scale will display Put 10 proceed to step 5 2 Press No to use the stored APW and proceed to step 5 3 Press Yes to est...

Page 12: ...limit Place the item to be weighed on the scale and press TARE Remove a portion of the item until it is within the ACCEPT range Zero Check Zero check is used when comparing subsequent samples to an initial reference sample In this case the UNDER limit must be a negative value and the OVER limit must be a positive value Place the reference item on the scale and press TARE Remove the reference sampl...

Page 13: ...by pressing the Function key rEAdY Ready will then be displayed Note If Set0 has been selected the countdown timer is not displayed The first weight larger than 5d will be displayed and held Automatic operation DYNAMIC is set to AUTOMATIC in the Mode menu 1 Press and hold Mode until dYNAmM Dynamic is displayed rEAdY Ready will then be displayed 2 Place the load on the pan to start measurement Duri...

Page 14: ...s removed With ACCUMULATE set to AUTO place the item on the scale The displayed value is accumulated automatically The icon will keep flashing until the weight is removed 3 8 2 Viewing and Clearing Statistical Data When the pan is cleared press the M key to display the statistical information To clear the accumulation data press the Tare key while the statistical information is displayed The displ...

Page 15: ... Put the weight 0 04 kg on the pan and press M key The Σ icon indicator will keep flashing until the weight is removed Step ii Remove the weight from the pan Put another weight 0 03 kg and press M key Step iii Remove the weight from the pan Step iv Press M key to display the statistical information Step v If required follow section 3 8 2 to clear the accumulation data ...

Page 16: ...ow opening Install the hook accessory into the access hole at the bottom of the scale as shown Do not over tighten tighten finger tight Mount the scale onto an appropriate assembly that allows free working space below the hook See figure 3 5 Note Never allow the scale to rest directly on the hook Figure 3 5 Setup for the Weigh Below Hook Weigh Below Hook Access ...

Page 17: ... End U n i t P r i n t C O M L O C k E n d Menu Items Reset kg g lb oz lb oz End Reset Stable A Print Contnt Layout Data Tr End Reset Baud Parity Stop Handsh Alt Cm End L Cal L Setup L Read L Mode L Unit L Print L COM End Notes Some modes units may not be available in all models When LEGAL FOR TRADE is set to ON Lock Switch in locked position the menu settings are affected as follows Calibration C...

Page 18: ...imits the current setting is displayed with all digits flashing Press the No button to begin editing The first digit is displayed flashing Press the No button to increment the digit or press the Yes button to accept the digit and move to the next digit The first digit can be used to scroll between positive and negative value Repeat this process for all digits Press the Yes button when the last dig...

Page 19: ...pan Press Yes to accept the weight or No to select an alternate weight C will be displayed while the reading is stored The display will show done if the calibration was successful The scale returns to the previous application mode and is ready for use TABLE 4 1 Suggested Span Calibration Mass sold separately Max Mass 1 Max Mass 1 1500g 1 5kg 3lb 15000g 15kg 30lb 3000g 3kg 6lb 30000g 30kg 60lb 6000...

Page 20: ...Tare is disabled ON the first stable gross weight is tared ON ACC stable gross loads within the accept limits are tared in Check weighing mode Beeper Signal Bp sIG Set how the beeper responds in Check Weighing mode OFF the beeper is disabled ACCEPT the beeper will sound when the weight is within the Accept range UNDER the beeper will sound when the weight is below the Under setting OVER the beeper...

Page 21: ...u to factory defaults NO not reset YES reset Stable Range Stable Set the amount the reading can vary while the stability symbol remains on 0 5d 0 5 scale division 1d 1 scale division 2d 2 scale division 5d 5 scale division Filter FILtEr Set the amount of signal filtering LOW less stability faster stabilization time MED normal stability stabilization time HI greater stability slower stabilization t...

Page 22: ... been exceeded Light LIGHt Set the light intensity LO low light intensity MED medium light intensity HI high light intensity Sleep SLEEP Set the display shut off time OFF disabled display will stay on until power off ON display will turn off in one minute in the absence of key or weight action Auto off A OFF Set the automatic shut off functionality OFF disabled 1 powers off after 1 minute of no ac...

Page 23: ...d ON enabled Check CHECk Set the sub mode OFF disabled WEIGH weight mode PERCNT percent mode COUNT count mode Dynamic DYNAmM Set the status OFF disabled MAN averaging and resetting are initiated manually SEMI averaging is automatically initiated resetting is manually initiated AUTO averaging and resetting are automatically initiated End Mode End Advance to the next menu or return to the top of the...

Page 24: ...r 2000 Series 4 7 Unit Menu This menu activates units so they will be accessible with the Units button The units in the menu must be turned on to be active Note Available units vary by model and local regulations ...

Page 25: ... each time the display is within the Checkweigh accept range and stability criteria are met CONT printing occurs continuously When ON STAB is selected set the condition for printing where LOAD printing occurs when the load is stable and greater than zero LOAD ZR printing occurs when any load is stable and equal to or greater than zero When INTER is selected set the Print Interval 1 to 3600 seconds...

Page 26: ...he status OFF disabled ON the Tare weight is printed Header Set the status OFF disabled ON the Header is printed Note See section 7 3 1 for instructions how to enter Header line Footer Set the status OFF disabled ON the Footer is printed Note See section 7 3 1 for instructions how to enter Footer line Mode Set the status OFF disabled ON the Mode is printed Unit Set the status OFF disabled ON the U...

Page 27: ...Start Menu in Windows XP and click Settings open Control Panel 2 Double click Accessibility Options in Control Panel 3 Select the General tab 4 Check Use Serial Keys and click the Settings button 5 Select the Serial Port set the Baud rate to 9600 and click OK 6 Close the Control Panel 7 Run Excel to open one blank sheet Focus the cursor on one item At this time if scale sends data to the PC throug...

Page 28: ...paritY Set the data bits and parity 7 EVEN 7 data bits even parity 7 ODD 7 data bits odd parity 7 NONE 7 data bits no parity 8 NONE 8 data bits no parity Stop bit stOP Set the number of stop bits 1 1 stop bit 2 2 stop bits Handshake HaNdSH Set the flow control method NONE no handshaking ON OFF XON XOFF software handshaking Reset no yes Baud Rate 300 9600 19200 Parity 7 even 7 odd 7 none 8 none Sto...

Page 29: ...he Print Tare and Zero commands Print Set the alternate command character for Print A to Z Tare Set the alternate command character for Tare A to Z Zero Set the alternate command character for Zero A to Z End COM End Advance to the next menu or return to the top of the current menu ...

Page 30: ...etup menu is not locked ON Setup menu is locked Lock Read l rEAD Set the status OFF Readout menu is not locked ON Readout menu is locked Lock Mode l mMod Set the status OFF Mode menu is not locked ON Mode menu is locked Lock Unit l Unit Set the status OFF Unit menu is not locked ON Unit menu is locked Lock Print l Print Set the status OFF Print menu is not locked ON Print menu is locked Reset no y...

Page 31: ... OFF COM menu is not locked ON COM menu is locked End Lock End Advance to the next menu or return to the top of the current menu 4 11 End Menu Press Yes to exit the menu and return to the current application mode Press No to advance to the Calibration menu ...

Page 32: ...11 42 19 43 26 17 16 16 15 15 14 14 13 13 12 12 43 26 44 32 17 17 16 16 15 15 14 14 13 13 12 44 32 45 38 18 17 17 16 16 15 15 14 14 13 13 45 38 46 45 18 18 17 17 16 16 15 15 14 14 13 46 45 47 51 19 18 18 17 17 16 16 15 15 14 14 47 51 48 58 19 19 18 18 17 17 16 16 15 15 14 48 58 50 06 20 19 19 18 18 17 17 16 16 15 15 50 06 51 13 20 20 19 19 18 18 17 17 16 16 15 51 13 52 22 21 20 20 19 19 18 18 17 1...

Page 33: ...ble Zero Range setting is locked at 2 Stable Range setting is locked at 1d Auto Zero Tracking setting is locked at 0 5d Units are locked at their current settings Stable Only is locked On Auto Print Continuous is disabled lb oz is locked Off 5 1 Settings Before verification and sealing perform the following steps 1 Verify that the menu settings meet the local weights and measures regulations 2 Per...

Page 34: ...d charge the battery Err 8 1 Power On Error Weight reading exceeds Power On Zero limit Err 8 2 Power On Error Weight reading below Power On Zero limit Err 8 3 Over Range Error Weight reading exceeds Overload limit Err 8 4 Under Range Error Weight reading below Underload limit Err 8 5 Tare out of range Adjust tare value to be within range Err 8 6 Display overflow Weight exceeds 6 digits Err 9 5 Cal...

Page 35: ...echnical support in the United States call toll free 1 800 526 0659 between 8 00 AM and 5 00 PM EST An OHAUS product service specialist will be available to provide assistance Outside the USA please visit our web site www ohaus com to locate the OHAUS office nearest you 6 4 Accessories In Use Cover 30037450 Weighing Below Hook 30025993 Printer Contact Ohaus Communication Cables Contact Ohaus ...

Page 36: ...005 lb 240oz x 0 01oz 15kg x 0 0005kg 15000g x 0 5g 30 lb x 0 001 lb 480oz x 0 02oz 30kg x 0 001kg 30000g x 1g 60 lb x 0 002 lb 960oz x 0 04oz 30 kg x 0 0001kg 30000g x 0 1g 60 lb x 0 0002 lb 960oz x 0 004oz Maximum Displayed Resolution 1 30000 1 300000 Repeatability 0 1 g 0 2 g 0 4 g 1 g 2 g 1 g Linearity 0 1 g 0 2 g 0 4 g 1 g 2 g 1 g Weighing Units g kg lb oz lb oz Tare Range To capacity by subt...

Page 37: ...Ranger 2000 Series EN 35 7 2 Drawings Figure 7 1 Dimensions 311 mm 12 2 in 327 mm 12 9 in 117 mm 4 6 in ...

Page 38: ...to print OFF Z Same as pressing Zero Key T Same as pressing Tare Key xT Download Tare value in grams positive values only Sending 0T clears tare if allowed PU Print current unit g kg lb oz lb oz t c custom xU Set scale to unit x 1 g 2 kg 3 lb 4 oz 5 lb oz 6 t xM Set scale to mode x M will scroll to next enabled mode PV Version print name software revision and LFT ON if LFT is set ON H x text Enter...

Page 39: ...ion line If Printx Content Info is ON MODE WEIGH Mode line If Printx Content Application Mode is ON Into is ON Weigh Mode with Accumulation Printout Description Comment 12 34 kg Result line If Printx Content Result is ON 12 34 kg G Gross value line If Printx Content Gross is ON and a tare value is entered 11 11 kg N Net value line If Printx Content Net is ON and a tare value is entered 1 23 kg T T...

Page 40: ...s entered 11 11 kg N Net value line If Printx Content Net is ON and a tare value is entered 1 23 kg T Tare value line If Printx Content Tare is ON and a tare value is entered MODE COUNT Mode line If Printx Content Application Mode is ON APW 5 g Information line If Printx Content Info is ON Check Weigh Mode Printout Description Comment 12 34 kg OVER Result line If Printx Content Result is ON UNDER ...

Page 41: ...ight aligned Space Application Status Right aligned Term Length 11 1 5 1 1 1 2 1 6 2 Non Check Weighing application Field Weight Right aligned Space Unit Right aligned Space Stability Space T N G PT Right aligned Term Length 11 1 5 1 1 1 2 2 Each field is followed by a single delimiting space ASCII 32 Definitions Weight Up to 11 characters right justified at immediate left of most significant char...

Page 42: ...s been damaged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other than OHAUS In lieu of a properly returned warranty registration card the warranty period shall begin on the date of shipment to the authorized dealer No other express or implied warranty is given by OHAUS...

Page 43: ......

Page 44: ... Avec des bureaux dans le monde entier Weltweite Geshäftsstellen Con uffici in tutto il mondo www ohaus com 30037458 P N 30037458D 2017 OHAUS Corporation all rights reserved todos los derechos reservados tous droits réservés Alle Rechte vorbehalten tutti i diritti riservati Printed in China Impreso en la China Impriméen Chine Gedruckt in China Stampato in Cina ...

Reviews: