Model S-100 Floor Scale 

 

 
 

Operation Manual 

 

 
 
 
 
 
 

Manual: xxxxxx   Issue:xx

Summary of Contents for S-100

Page 1: ...Model S 100 Floor Scale Operation Manual Manual xxxxxx Issue xx...

Page 2: ...DOUBT SEEK ADVICE FROM A QUALIFIED ELECTRICIAN Routine maintenance To avoid the possibility of electric shock or damage to the machine always switch off the machine and isolate from the power supply...

Page 3: ...nu 14 Service Configuration Settings 15 Technical Specifications 16 Error Messages 16 Declarations of Compliance 17 Contact Information Technical Support 17 Product Identification Table Part Number Ca...

Page 4: ...oadcell socket Power Adapter Socket Power Supply Connecting the Indicator to the Platform 6VDC 500mA 230V 50Hz Wire the cable attached to the base as shown if necessary 4 2 3 1 1 RED 2 RED 3 RED 4 RED...

Page 5: ...Indicator and Platform Assembly Rem ove packagin g pa ds x2 x2 1 2 3 x4 4 5...

Page 6: ...0 Weight is zero On Off Zero On Off Zero kg lb Unit of measure Flashing LED means that the weight reading is not stable Units kg lb Unit of measurement NET Tare is activated Hold Print Hold Print CH B...

Page 7: ...tion Turning On and Zeroing the Scale On Off Zero ZERO 0 Turning off the scale On Off Zero Selecting Unit of Measure Units kg lb Using the Tare 1 2 3 NET 4 NET Removing the Tare On Off Zero 0 ZERO kg...

Page 8: ...old function with manual release This function needs to be setup in Parameter P3 1 shown on page 10 1 Hold Print Hold 2 3 Removing Hold Hold Print Hold Print Function For communications to a printer o...

Page 9: ...ecting parameter Tare Changing data within the parameter Units Saving data Tare Exiting setup On Off Zero ZERO 0 Gross 45 00kg Tare 1 35kg Net 43 65kg Hold Print kg Moves flashing digit Units Change f...

Page 10: ...ctable hold window from 3 to 50 divisions Will hold display reading once stable within a selectable weight range to release the hold button must be represses P3 0 P3 1 P3 2 P3 3 to 50 P4 x RS232 Seria...

Page 11: ...tare LF H1H2H3 CR ETX U CR Changes the units LF u1u2 CR LF H1H2H3 CR ETX L CR Activates the hold function LF H1H2H3 CR ETX X CR Switches off the scale Indicator switches off Unrecognised command LF CR...

Page 12: ...minus decimal point unit CR EXT LF Tare reading decimal point unit CR EXT LF Net reading minus decimal point unit CR EXT RS232 serial interface wiring DE 9 Female Scale DE 9 Male Host Pin Name Direct...

Page 13: ...bration weight 0 KG 1 LB Factory Calibration Settings shown in bold Primary Capacity Displayed Resolution Units Selectable Capacity P7 P8 P9 P10 150 lb x 0 02 lb 7500 75 kg x 0 01kg 16 1 2 1 150 x 0 0...

Page 14: ...g lb 5 Single point calibration enter the same weight in again and move to number 7 For 2 point calibration enter in the second calibration weight between 25 100 full capacity Tare 6 7 Tare 8 9 Tare Z...

Page 15: ...5 50 6 100 7 no limitation P13 00 15 3 Zero button range Full scale 00 1 06 20 12 5 01 2 07 no limit 13 10 02 3 08 1 14 20 03 4 09 2 25 no limit 04 5 10 3 05 10 11 4 P14 0 1 2 0 Signal within power on...

Page 16: ...ill be turned off in two minutes automatically AC Adapter 10 5 12VDC 600mA with central positive 5 7 VDC 9 VDC Error Messages Error Message Definition Required Solution Troubleshooting 0 Weight above...

Page 17: ...ications Le pr sent appareil num rique n met pas de bruits radio lectriques d passant les limites applicables aux appareils num riques de la Classe A prescrites dans le R glement sur le brouillage rad...