Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
2-17
Reference Air Set, Solenoids, and Blowback Option with
COe Zero Function
Figure 2-13 shows the piping arrangement for the OCX 8800 with the
blowback and autocalibration options when the COe Zero Function is used.
The arrangement is similar to Figure 2-12 except instrument air is used as the
Hi O
2
test gas. Refer to Sectio 3 for details of the function.
Reference Air Set, Solenoids, and Blowback Option without
COe Zero Function
Installing an OCX 8800 with the blowback option requires the addition of air
operated blowback valve, regulator and gage, and check valve.
Figure 2-12 shows the piping arrangement for the OCX 8800 with the
blowback and autocalibration options. Figure 2-14 shows the piping
arrangement for the OCX 8800 with the blowback option, but without
autocalibration (without test gas solenoids).
When the reference air set, calibration gas solenoids, and blowback options
are included with your transmitter, use the following procedure to install the
pneumatic system components.
1. Connect the calibration gas sources according to the instructions in the
previous paragraph “Reference Air Set and Solenoids Option”, steps 2
through 5.
2. Connect a clean, dry, instrument-quality supply of air (20.95% O
2
) to the
35 psig and 55 psig pressure regulators. The inlet to the 35 psig
regulator accepts a 1/8" NPT fitting. The inlet to the 55 psig regulator
accepts a 1/4" NPT fitting.
3. See the upper leg of the instrument air supply. Connect the output of the
35 psi regulator/filter to one port of the normally-closed air-operated
solenoid valve, and to the inlet side of the dilution air flow meter.
4. Connect the dilution air flow meter output to the DILUTION AIR inlet
fitting on the sensor housing.
5. Install an instrument air line between the open port of the normally-open
air-operated solenoid valve and the tee fitting on the sensor housing.
6. Connect the output of the 55 psi regulator/filter to one port of the
normally-open air-operated solenoid valve, and to the instrument air
inlet on the back of the electronics housing.
7. Install an air line between the open port of the normally-closed
air-operated solenoid valve and the check valve inlet fitting on the
sensor housing.
8. Install an air line between the instrument air outlet fitting on the
electronics housing and the control air inlet fitting on the air-operated
solenoid valve.
Содержание OCX 8800
Страница 2: ......
Страница 4: ......
Страница 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Страница 22: ...1 10 Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 Figure 1 7 Optional Sample Tube Support ...
Страница 33: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 2 5 Figure 2 3 Adapter Plate Installation ...
Страница 70: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 6 Figure 4 4 LOI Menu Tree Sheet 3 of 4 ...
Страница 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Страница 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Страница 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Страница 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Страница 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Страница 185: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 9 Figure 9 6 Remote Electronics Housing Cable Connections ...
Страница 209: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 33 Figure 9 26 Electronics Housing Cable Connections ...
Страница 213: ......
Страница 217: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 5 Figure 10 2 Sensor Housing Components Sheet 2 of 2 ...
Страница 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Страница 225: ......
Страница 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...