COe PURGE / ZERO
FEATURE
This feature provides a way to periodically flood the COe sensor with air to
perform two functions; a) provide additional oxygen to help burn off any
combustible residue from the COe sensor, and b) allow for optional
adjustment of the COe calibration constant. If the OCX8800 is configured to
update the COe calibration constant, only the constant is updated. The COe
calibration slope is not affected. To update both the constant and slope a full
calibration must take place.
The feature uses the calibration solenoid that is also used for high O2 test gas
and COe zero gas. For the feature to work properly, instrument air is used as
the high O2 test gas. This also requires the high O2 test gas value to be set
at 20.95%.
As an option, a two way valve may be installed to switch the high O2 test gas
between the normal calibration gas and instrument air. This allows the OCX
8800 to use a specified calibration gas for calibration, then instrument air for
the COe zero feature. Switching between the two gases must be manually
coordinated between scheduled calibrations and COe zero events.
When the COe zero feature is used, special pneumatic connections are
required. See Section 2, Pneumatic Installation, for details.
The COe zero feature is only valid if the OCX 8800 is supplied with calibration
solenoids and the solenoids have been activated.
An OCX 8800 shipped from the factory must be configured before the COe
zero feature can be implemented. This same process must be performed any
time a replacement card stack is installed.
Configuring the COe Zero Feature with the Field Communicator - HART
1. Use the 375 Field Communicator or AMS software to access the HART
menu.
2. From the
DETAILED SETUP
menu, select
INPUT/OUTPUT
.
3. From the
INPUT/OUTPUT
menu, select
COE ZERO
.
4. From the
COE ZERO
menu, select the functions as follows:
Zero Enabled
- Select "Yes" or "No" to enable or disable this feature.
Zero Intrvl
- Length of time between COe zero events. Range is 60 to
480 minutes. Default is 60 minutes.
During the COe Zero Function the analog output signals may track the oxygen and
combustibles readings if configured to do so. To avoid a potentially dangerous operating
condition, the OCX 8800 must be removed from the automatic combustion control loop
before performing the COe Zero Function procedure.
At the completion of the COe Zero Function, the COe analog output signal will change if the
Zero Update parameter is set to "Yes".
Содержание OCX 8800
Страница 2: ......
Страница 4: ......
Страница 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Страница 22: ...1 10 Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 Figure 1 7 Optional Sample Tube Support ...
Страница 33: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 2 5 Figure 2 3 Adapter Plate Installation ...
Страница 70: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 6 Figure 4 4 LOI Menu Tree Sheet 3 of 4 ...
Страница 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Страница 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Страница 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Страница 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Страница 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Страница 185: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 9 Figure 9 6 Remote Electronics Housing Cable Connections ...
Страница 209: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 33 Figure 9 26 Electronics Housing Cable Connections ...
Страница 213: ......
Страница 217: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 5 Figure 10 2 Sensor Housing Components Sheet 2 of 2 ...
Страница 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Страница 225: ......
Страница 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...