Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
5-10
Manual O
2
/COe Calibration using the Field Communicator - Fieldbus
To perform a manual O
2
/COe calibration using the 375 Field Communicator,
use the following procedure. If necessary, refer to Section 6, Field
Communicator, for the Fieldbus menu tree.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
1. To calibrate from Fieldbus, the first step is to set the Transducer Block to
Out of Service Mode (OOS).
2. To set the OOS mode, select
Transducer
then select
Process
, followed
by select
Out of Service
in the
Target Mode
.
3. From
Transducer
screen select
Methods
.
4. Select
OCX Calibration
from the
Methods
menu.
5. From the
OCX Calibration
screen, select
Calibrate O2 Sensor
for O2
calibration, select
Calibrate Combustibles Sensor
for COe calibration,
or select
Calibrate Both Sensors
if calibrate both sensors is desired.
6. In the calibration screen, a
Loop should be removed from automatic
control
warning appears. Remove the device from any automatic
control loops to avoid a potentially dangerous operating condition and
press
OK
.
7. The
Select Action
screen should look like the following. From the
Select Action screen, select
Update Display
to refresh the calibration
status, select
Next Calibration Step
to continue calibration, select
Abort Calibration
to abort calibration or
Exit
to exit calibration. Select
one item from the list and press
OK
. The Select Action screen is static
and data will not be automatically refreshed.
Calibration Step = Idle
Step Time Remaining = 0 seconds
O2 Value = 0.40 %
Combustibles Value = 1000 ppm
1. Update Display
2. Next Calibration Step
3. Abort Calibration
4. Exit
8. The
Calibration Screen
should look like the following, press
OK
to
continue. The
Calibration Screen
should be automatically refreshed,
however it may take a while for the data to refresh.
Calibration Step = Apply O2 Low Gas
Step Time Remaining = 0 seconds
O2 Value = 0.40 %
Combustibles Value = 1000 ppm
Press Next for Selection
Содержание OCX 8800
Страница 2: ......
Страница 4: ......
Страница 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Страница 22: ...1 10 Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 Figure 1 7 Optional Sample Tube Support ...
Страница 33: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 2 5 Figure 2 3 Adapter Plate Installation ...
Страница 70: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 6 Figure 4 4 LOI Menu Tree Sheet 3 of 4 ...
Страница 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Страница 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Страница 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Страница 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Страница 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Страница 185: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 9 Figure 9 6 Remote Electronics Housing Cable Connections ...
Страница 209: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 33 Figure 9 26 Electronics Housing Cable Connections ...
Страница 213: ......
Страница 217: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 5 Figure 10 2 Sensor Housing Components Sheet 2 of 2 ...
Страница 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Страница 225: ......
Страница 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...