Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
7-22
Transducer Block
Channel Status
The status of channel 1 to 4 are affected by the state of unit alarm. In all
cases, the channel will read what it believes the correct sensor values.
Self-Clearing alarms are reset when the alarm condition goes away. All others
require the device be restarted.
Table 7-18 indicates channel status under indicated device conditions
(Occurence). Unless otherwise indicated in the table, the status values are:
• Occurence: Normal
• Quality Value: GOOD
• Quality Substatus Value: NON_SPECIFIC
• Limit Value: NOT_LIMITED
Table 7-18. I/O Channel Status
Channel
Occurence
Quality Value
Quality Substatus Value
Limit Value
1,2
Powerup, Warmup, Stablize
BAD
NON_SPECIFIC
NOT_LIMITED
1,2
Normal
GOOD
NON_SPECIFIC
NOT_LIMITED
1,2
Calibrating, Cal Verify, Blow Back,
COe Zero,
UNCERTAIN
SENSOR_CONVERSION_INACCURATE
NOT_LIMITED
1,2
Alarm
(Temperature Low & High, Cell
Bad/Error)
UNCERTAIN
SENSOR_CONVERSION_INACCURATE
NOT_LIMITED
1,2
System Fault
BAD
DEVICE_FAILURE
NOT_LIMITED
3,4
Powerup, Warmup, Stablize
GOOD
NON_SPECIFIC
NOT_LIMITED
3,4
Normal
GOOD
NON_SPECIFIC
NOT_LIMITED
3,4
Calibrating, Cal Verify, Blow Back
GOOD
NON_SPECIFIC
NOT_LIMITED
3,4
Alarm, System Fault
(Temperature Related Alarms: T/C
Open, T/C Shorted, T/C Reversed,
ADC Error)
BAD
DEVICE_FAILURE
CONSTANT
3,4
Temp Low & High
Good
ACTIVE_BLOCK_ALARM
NOT_LIMITED
Transducer Block
Simulate
Support Transducer
Block Errors
Setting PWA_SIMULATE to ON also allows simulating TB status and to check
the correct mapping onto the
PWA’s FAILED_ACTIVE, MAINT_ACTIVE, and
ADVISE_ACTIVE parameters.
Out of Service
Set whenever the transducer block actual mode is OOS.
Input Failure
Set whenever there is a communication error between the Fieldbus A2D card
and the OCX.
Simulation Active
Set whenever the Fieldbus Simulate Switch is set to ON at the Fieldbus A2D
card or software simulate option is enabled.
Other Error
Set whenever XD_ERROR is non-zero.
Содержание OCX 8800
Страница 2: ......
Страница 4: ......
Страница 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Страница 22: ...1 10 Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 Figure 1 7 Optional Sample Tube Support ...
Страница 33: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 2 5 Figure 2 3 Adapter Plate Installation ...
Страница 70: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 6 Figure 4 4 LOI Menu Tree Sheet 3 of 4 ...
Страница 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Страница 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Страница 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Страница 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Страница 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Страница 185: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 9 Figure 9 6 Remote Electronics Housing Cable Connections ...
Страница 209: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 33 Figure 9 26 Electronics Housing Cable Connections ...
Страница 213: ......
Страница 217: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 5 Figure 10 2 Sensor Housing Components Sheet 2 of 2 ...
Страница 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Страница 225: ......
Страница 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...