Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
5-11
9. When the calibration status is at the
Apply O2 Low Gas/ Apply Comb
Low Gas
step, switch on O2 Low Gas/COe Low Gas. Verify the O2
concentration measured matches the
O2 Low Gas
parameter in the
setup. The unit samples reference air as the COe Low Gas. Press
OK
when ready.
10. Select
Next Calibration Step
and press
OK
to start applying the test
gas. The time to apply the test gas is specified by the
Gas Time
.
11. The calibration step will change to
Flow O2 Low Gas/ Flow Comb Low
Gas
and then
Read O2 Low Gas/ Read Comb Low Gas
for a period of
time.
12. When ready, Calibration Step will stop at the
Apply O2 High Gas/
Apply Comb High Gas
. Switch off the O2 Low Gas/Sample Reference
Air and switch on the O2 High Gas/Comb Test Gas. Verify the O2/COe
concentration measured matches the
O2 High Gas/COe Test Gas
parameter in the setup. Press
OK
when ready.
13. Select
Next Calibration Step
to start applying the test gas. The time to
apply the test gas is specified by the
Gas Time
.
14. The calibration step will change to
Flow O2 High Gas/ Flow Comb
High Gas
and then
Read O2 High Gas/ Read Comb High Gas
for a
period of time.
15. Skip over to Step 19,
Stop Gas
, if executing
Calibrate O2 Sensor
or
Calibrate Combustibles Sensor
.
16. The calibration step will change to
Read Comb Low Gas
for a period of
time. When ready, Calibration Step will stop at
Apply Comb High Gas
.
Switch off the O2 High Gas and switch on the COe Test Gas. Verify the
COe concentration measured matches the
COe Test Gas
parameter in
the setup.
17. Select
Next Calibration Step
to start applying the test gas. The time to
apply the test gas is specified by the
Gas Time
.
18. The calibration step will change to
Flow Comb High Gas
and then
Read Comb High Gas
for a period of time.
19. When ready, Calibration step will stop at
Stop Gas
. Switch off the O2
High Gas/COe Test Gas, Press
OK
when ready. Select Next Calibration
Step to start purging gas. The time to purge gas is specified by the
Purge Time
.
20. When the
Purge
step is complete, the Calibration Step will be at
Idle
.
The Calibration Failed alarm will be set if the calibration has failed.
21. When calibration is complete. Select
Exit
to exit the calibration method.
Содержание OCX 8800
Страница 2: ......
Страница 4: ......
Страница 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Страница 22: ...1 10 Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 Figure 1 7 Optional Sample Tube Support ...
Страница 33: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 2 5 Figure 2 3 Adapter Plate Installation ...
Страница 70: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 6 Figure 4 4 LOI Menu Tree Sheet 3 of 4 ...
Страница 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Страница 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Страница 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Страница 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Страница 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Страница 185: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 9 Figure 9 6 Remote Electronics Housing Cable Connections ...
Страница 209: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 33 Figure 9 26 Electronics Housing Cable Connections ...
Страница 213: ......
Страница 217: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 5 Figure 10 2 Sensor Housing Components Sheet 2 of 2 ...
Страница 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Страница 225: ......
Страница 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...