Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
7-34
Troubleshooting
Refer to Table 7-25 to troubleshoot any problems that you encounter.
Table 7-25. Troubleshooting
Mode will not
leave OOS
Process and/or
block alarms will
not work
Value of output
does not make
sense
1. Target mode not set
2. Configuration error
3. Resource block
4. Schedule
1. Features
2. Notification
3. Status Options
1. Linearization Type
2. Scaling
1. Set target mode to something other than OOS.
2. BLOCK_ERR will show the configuration error
bit set. The following are parameters that must be
set before the block is allowed out of OOS:
a. CHANNEL must be set to a valid value and
cannot be left at initial value of 0.
b. XD_SCALE.UNITS_INDX must match the
units in the transducer block channel value.
c. L_TYPE must be set to Direct, Indirect, or In-
direct Square Root and cannot be left at ini-
tial value of 0.
3. The actual mode of the Resource block is
OOS. See Resource Block Diagnostics for cor-
rective action.
4. Block is not scheduled and therefore cannot
execute to go to Target Mode. Schedule the block
to execute.
1. FEATURES_SEL does not have Alerts en-
abled. Enable the Alerts bit.
2. LIM_NOTIFY is not high enough. Set equal to
MAX_NOTIFY.
3. STATUS_OPTS has Propagate Fault Forward
bit set. This should be cleared to cause an alarm
to occur.
1. Does not make L_TYPE must be set to Direct,
Indirect, or Indirect Square Root and cannot be
left at initial value 0.
2. Scaling parameters are set incorrectly:
a. XD_SCALE.EU0 and EU100 should match
that of the transducer block channel value.
b. OUT_SCALE.EU0 and EU100 are not set
properly.
Cannot Set
HI_LIMIT,
HI_HI_LIMIT
LO_LIMIT, or
LO_LO_LIMIT
Values
1. Scaling
1. Limit values are outside the OUT_SCALE.EU0
and OUT_SCALE.EU100 values. Change
OUT_SCALE or set values within range.
Symptom
Possible Cause
Corrective Action
Содержание OCX 8800
Страница 2: ......
Страница 4: ......
Страница 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Страница 22: ...1 10 Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 Figure 1 7 Optional Sample Tube Support ...
Страница 33: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 2 5 Figure 2 3 Adapter Plate Installation ...
Страница 70: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 6 Figure 4 4 LOI Menu Tree Sheet 3 of 4 ...
Страница 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Страница 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Страница 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Страница 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Страница 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Страница 185: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 9 Figure 9 6 Remote Electronics Housing Cable Connections ...
Страница 209: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 33 Figure 9 26 Electronics Housing Cable Connections ...
Страница 213: ......
Страница 217: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 5 Figure 10 2 Sensor Housing Components Sheet 2 of 2 ...
Страница 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Страница 225: ......
Страница 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...