Alerts
Alarms
AMS
Tab
What does the
alert indicate?
Recommended
Action
Description
NV Memory
Transducer
FF/
The non-volatile
At startup, wait 2
This alarm will generally occur during a startup
Failure
Block:
Device
parameter
minutes with power
condition. Rarely, the device could be powered
EEPROM
Alerts
storage on the
applied and then
down while a parameter is being stored to the
Corrupt
CPU board has
cycle power again or non-volatile memory. The parameter will then be
become
reset device.
tagged as bad on the next power on and the
Resource Block:
unreliable.
memory contents will be written with default
1) Manufacturing
parameters. Calibration data may be lost and the
2) Block Integrity
unit should be recalibrated. If the unit does not
Error
recover automatically, the memory may be faulty
3) NV Integrity
and the electronics stack should be replaced.
Error
Refer to Section 8: Troubleshooting for details.
4) ROM Integrity
Error
NV Writes
Resource Block: FF/
A high number of writes has been detected to
Deferred
1) NV Writes
Device
non-volatile memory. To prevent premature failure
Deferred
Alerts
of the memory, the write operations have been
deferred. The data will be saved on a 6 hour
cycle. This condition usually exists because a
program has been written that writes to Function
block parameters not normally expected to be
written to on a cyclic basis.
High Electronics Transducer
Device
This alert
Evaluate mounting
The device may require special mounting
Temp-
Block:
Alerts
indicates that the location and
considerations if installed in a very hot location.
erature
electronics
environment.
High Electronic
Temperature
temperature has
exceeded 80
°
C.
The device will
cease to operate
reliably beyond
85
°
C.
ADC Failure
Transducer
Device
This alert
Cycle power or
The Analog to Digital Converter (ADC) is
Block:
Alerts
indicates faulty
replace the
continuously monitored by the device for correct
1) ADC Timeout
operation of the
electronics stack.
operation. Refer to Section 8: Troubleshooting for
Error
device
details.
2) ADC
electronics.
Reference
Error
Содержание OCX 8800
Страница 2: ......
Страница 4: ......
Страница 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Страница 22: ...1 10 Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 Figure 1 7 Optional Sample Tube Support ...
Страница 33: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 2 5 Figure 2 3 Adapter Plate Installation ...
Страница 70: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 6 Figure 4 4 LOI Menu Tree Sheet 3 of 4 ...
Страница 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Страница 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Страница 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Страница 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Страница 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Страница 185: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 9 Figure 9 6 Remote Electronics Housing Cable Connections ...
Страница 209: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 33 Figure 9 26 Electronics Housing Cable Connections ...
Страница 213: ......
Страница 217: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 5 Figure 10 2 Sensor Housing Components Sheet 2 of 2 ...
Страница 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Страница 225: ......
Страница 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...