Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
3-6
Configuring the Calibration Solenoids with the LOI
1. Use the "Z" pattern to enter the LOI menu tree.
2. From the
SYSTEM
menu, select
Calib Setup
.
3. From the
Calib Setup
menu, select
Use Solenoids
. Select
Yes
to
enable the solenoids.
BLOWBACK FEATURE
The blowback feature blows instrument air back through the center of the
internal filter and out the sample tube of the probe. This removes built up dirt
and particulate from the internal filter, sample line and any optional in-situ filter
on the end of the sample tube. The blowback feature is normally used in
systems that have heavy particulate in the process stream.
The blowback feature requires the optional blowback hardware to be properly
installed external to the OCX 8800. See Section 2, Pneumatic Installation, for
details.
An OCX 8800 shipped from the factory must be configured before blowback
can be implemented. This same process must be performed any time a
replacement card stack is installed.
Configuring Blowback with the Field Communicator - HART
1. Use the 375 Field Communicator or AMS software to access the HART
menu.
2. From the
DETAILED SETUP
menu, select
INPUT/OUTPUT
.
3. From the
INPUT/OUTPUT
menu, select
BLOWBACK
.
4. From the
BLOWBACK
, select
BlBk Enabled
. Select
Yes
to enable
blowback. Also set the following parameters:
BlBk Intrvl
- Length of time between blowback events (60 minutes
recommended).
BlBk Period
- Length of time blowback is activated (5 seconds
recommended).
BlBk Purge Time
- Length of time after blowback is complete before
oxygem/combustibles readings are considered valid (Set as required by
the application).
5. Manually initiate blowback from DIAG/SERVICE, then BLOW BACK,
When select BLOWBACK.Configuring Blowback with the Field
Communicator - Fieldbus
1. Use the 375 Field Communicator or AMS software to access the
Fieldbus menu.
2. From the
TRANSDUCER
block menu, select
Alarm Relay/Blowback
.
3. From the
Alarm Relay/Blowback
menu, select
Blowback
.
4. From the
Blowback
menu, select
Blowback Enabled
. Also set the
following parameters:
Blowback Interval
- Length of time between blowback events (60
minutes recommended).
Blowback Period
- Length of time blowback is activated (5 seconds
recommended).
Содержание OCX 8800
Страница 2: ......
Страница 4: ......
Страница 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Страница 22: ...1 10 Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 Figure 1 7 Optional Sample Tube Support ...
Страница 33: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 2 5 Figure 2 3 Adapter Plate Installation ...
Страница 70: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 6 Figure 4 4 LOI Menu Tree Sheet 3 of 4 ...
Страница 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Страница 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Страница 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Страница 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Страница 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Страница 185: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 9 Figure 9 6 Remote Electronics Housing Cable Connections ...
Страница 209: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 33 Figure 9 26 Electronics Housing Cable Connections ...
Страница 213: ......
Страница 217: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 5 Figure 10 2 Sensor Housing Components Sheet 2 of 2 ...
Страница 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Страница 225: ......
Страница 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...