Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
5-3
Autocalibration Setup using the Field Communicator - Fieldbus
NOTE
Automatic calibration is only available on units equipped with calibration
solenoids.
Use the following procedure to specify a time interval (in hours) at which the
OCX 8800 will automatically calibrate.
1. From the
Transducer
screen select
O2 Cal
or
COe Cal
.
2. From the
O2 Cal
or
COe Cal
screen, select
O2 Cal Setup
or
COe
Setup
.
3. If the unit is equipped with calibration solenoids and timed automatic
calibration are desired, select
Solenoids
, then select
Yes
. Select
No
to
disable the calibration solenoids.
4. Select
O2 Cal Interval
and/or
COe Cal Interval
and enter the desired
time in hours between automatic calibrations. To disable automatic
calibration for O2 and COe, enter 0 for both
Cal Interval
parameters.
5. If desired, the
O2 Next Cal
and
CO2 Next Cal
(next calibration time)
parameters can be changed to synchronize a calibration at a specific
day or time.
OPERATOR - INITIATED
AUTOCALIBRATION
An operator can initiate an automatic calibration at any time provided that the
unit is equipped with calibration solenoids.
Autocalibration using the optional LOI
To initiate a calibration using the LOI, perform the following steps on the LOI
menu tree. Refer to Section 4, Using the LOI, for the LOI menu tree.
1. From the
CALIBRATION
menu, use the right-pointing arrow to select
the
Cal Control
menu.
2. Select
Start Cal-O2
,
Start Cal COe
, or
Start Cal Both
to start the
calibration. Select
Cal Verify
to access the calibration window.
3. At the prompt, use the right-pointing arrow to initiate automatic
calibration.
Autocalibration using the Field Communicator - HART
To initiate an automatic calibration using 375 Field Communicator, perform
the following steps. Refer to Section 6, Field Communicator, for the HART
menu tree.
1. Select
DIAG/SERVICE
from
DEVICE SETUP
menu.
2. Select
CALIBRATION
from the
DIAG/SERVICE
menu.
3. Select
CAL CONTROL
from the
CALIBRATION
menu.
4. Select
CAL METHODS
from the
CAL CONTROL
menu.
5. From the
CAL METHODS
menu, select the type of calibration desired:
O2 Calibration
,
COe Calibration
, or
O2 and COe Calibration
.
Содержание OCX 8800
Страница 2: ......
Страница 4: ......
Страница 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Страница 22: ...1 10 Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 Figure 1 7 Optional Sample Tube Support ...
Страница 33: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 2 5 Figure 2 3 Adapter Plate Installation ...
Страница 70: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 6 Figure 4 4 LOI Menu Tree Sheet 3 of 4 ...
Страница 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Страница 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Страница 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Страница 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Страница 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Страница 185: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 9 Figure 9 6 Remote Electronics Housing Cable Connections ...
Страница 209: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 33 Figure 9 26 Electronics Housing Cable Connections ...
Страница 213: ......
Страница 217: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 5 Figure 10 2 Sensor Housing Components Sheet 2 of 2 ...
Страница 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Страница 225: ......
Страница 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...