Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
5-2
4. Scroll down to the last item
Use Solenoids
. If the unit is equipped with
calibration solenoids and timed automatic calibration is desired, select
Yes
.
5. Use the up-pointing key to select the item
O2 Out Tracks
. Select
Yes
or
No
to determine if updates to the O
2
lock value will take place.
6. Use the down-pointing key to select the item
COe Out Tracks
. Select
Yes
or
No
to define if updates to the COe lock value will take place.
7. Use the down-pointing key to select the item
O2 Cal Interval
. Enter the
amount of time in days and hours that is desired between automatic
calibrations.
8. Use the down-pointing key to select the next item
O2 Next Cal
. Enter
the amount of time in hours until the next automatic calibration. Select
the left-pointing key three times to move back to the LOI operating
display.
Autocalibration Setup using the Field Communicator - HART
NOTE
Automatic calibration is only available on units equipped with calibration
solenoids.
Use the following procedure to specify a time interval (in hours) at which the
OCX 8800 will automatically calibrate.
1. From the
DEVICE SETUP
screen, select
DETAILED SETUP
.
2. From the
DETAILED SETUP
screen, select
CAL SETUP
, then select
O2 CAL PARAMS
or
COE CAL PARAMS
.
3. If the unit is equipped with calibration solenoids and timed automatic
calibrations are desired, select
Solenoids
, then select
Y
es. Select
N
o to
disable the calibration solenoids.
4. Select
O2 Cal Intrvl
(O
2
calibration interval) and enter the desired time
in hours between automatic O
2
calibrations. Select
COE Cal Intrvl
and
enter the desired time between automatic COe calibrations. To disable
automatic calibration for O
2
and COe, enter 0 for both
Cal Intrvl
parameters.
5. If desired, the
O2 Next Cal Time
and the
COe Next Cal Time
(next
calibration time) parameters can be changed to synchronize a
calibration at a specific day or time.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
When setting automatic calibration times, CalIntrvl and NxtCalTm should be set so that O
2
and COe are NOT calibrated simultaneously.
Содержание OCX 8800
Страница 2: ......
Страница 4: ......
Страница 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Страница 22: ...1 10 Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 Figure 1 7 Optional Sample Tube Support ...
Страница 33: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 2 5 Figure 2 3 Adapter Plate Installation ...
Страница 70: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 6 Figure 4 4 LOI Menu Tree Sheet 3 of 4 ...
Страница 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Страница 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Страница 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Страница 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Страница 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Страница 185: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 9 Figure 9 6 Remote Electronics Housing Cable Connections ...
Страница 209: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 9 33 Figure 9 26 Electronics Housing Cable Connections ...
Страница 213: ......
Страница 217: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 5 Figure 10 2 Sensor Housing Components Sheet 2 of 2 ...
Страница 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Страница 225: ......
Страница 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...