![Ronde & Schwarz R&S FPL1000 Скачать руководство пользователя страница 765](http://html1.mh-extra.com/html/ronde-and-schwarz/rands-fpl1000/rands-fpl1000_user-manual_1484989765.webp)
Remote Commands
R&S
®
FPL1000
765
User Manual 1178.3370.02 ─ 09
[SENSe:]ADJust:CONFigure:DURation:MODE
<Mode>
In order to determine the ideal reference level, the R&S
FPL1000 performs a measure-
ment on the current input data. This command selects the way the R&S
FPL1000
determines the length of the measurement .
Parameters:
<Mode>
AUTO
The R&S
FPL1000 determines the measurement length auto-
matically according to the current input data.
MANual
The R&S
FPL1000 uses the measurement length defined by
[SENSe:]ADJust:CONFigure:DURation
*RST:
AUTO
Manual operation:
See
"Resetting the Automatic Measurement Time (Meastime
See
"Changing the Automatic Measurement Time (Meastime
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer
<Threshold>
When the reference level is adjusted automatically using the
on page 766 command, the internal attenuators and the preamplifier are also
adjusted. In order to avoid frequent adaptation due to small changes in the input signal,
you can define a hysteresis. This setting defines a lower threshold the signal must fall
below (compared to the last measurement) before the reference level is adapted auto-
matically.
Parameters:
<Threshold>
Range:
0 dB to 200 dB
*RST:
+1 dB
Default unit: dB
Example:
SENS:ADJ:CONF:HYST:LOW 2
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level falls below 18
dBm.
Manual operation:
See
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer
<Threshold>
When the reference level is adjusted automatically using the
on page 766 command, the internal attenuators and the preamplifier are also
adjusted. In order to avoid frequent adaptation due to small changes in the input signal,
you can define a hysteresis. This setting defines an upper threshold the signal must
exceed (compared to the last measurement) before the reference level is adapted
automatically.
Setting Basic Spectrum RF Measurement Parameters