The Spectrum Application (RF Measurements)
R&S
®
FPL1000
366
User Manual 1178.3370.02 ─ 09
tionally. Logarithmic averaging is thus fully taken into account, which yields a value
that is 1.45
dB lower than that of linear averaging.
●
To allow for a more stable noise display, eight trace points on each side of the mea-
surement frequency are averaged.
●
For span > 0, the measured values are averaged versus time (after a sweep).
The R&S
FPL1000 noise figure can be calculated from the measured power density
level. It is calculated by deducting the set RF attenuation (RF Att) from the displayed
noise level and adding 174 to the result.
The individual marker settings correspond to those defined in the "Marker" dialog box
(see
Chapter 7.9.2.1, "Individual Marker Setup"
on page 349). Any settings to the
marker state or type changed in the "Marker Function" dialog box are also changed in
the "Marker" dialog box and vice versa.
Remote commands:
"Example: Measuring Noise Density"
CALCulate<n>:MARKer<m>:FUNCtion:NOISe[:STATe]
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult?
Marker Usage