This data is stored to the C:\Users\Public\Documents\Rohde-Schwarz\Analyzer direc-
tory on the instrument.
The file name consists of the unique device ID and the current date and time of the file
creation.
If you contact the Rohde
&
Schwarz support to get help for a certain problem, send
these files to the support in order to identify and solve the problem faster.
Remote command:
Save Device Footprint
Creates an
*.xml
file with information on installed hardware, software, image and
FPGA versions. The
*.xml
file is stored under
C:\ProgramData\Rohde-Schwarz\ZNL-FPL\devicedata\xml\
on the instru-
ment. It is also included in the service ZIP file (see
"Create R&S Support Information"
9.5.2
Self-test Settings and Results
Access
: [Setup] > "Service" > "Selftest"
If the R&S
FPL1000 fails you can perform a self-test of the instrument to identify any
defective modules.
Once the self-test is started, all modules are checked consecutively and the test result
is displayed. You can abort a running test.
In case of failure a short description of the failed test, the defective module, the associ-
ated value range and the corresponding test results are indicated.
A running Sequencer process is aborted when you start a self-test.
If you start a self-test remotely, then select the "Local" softkey while the test is still run-
ning, the instrument only returns to the manual operation state after the test is comple-
ted. In this case, the self-test cannot be aborted.
Service Functions