Remote Commands
R&S
®
FPL1000
636
User Manual 1178.3370.02 ─ 09
To query the results of the 1
dB compression point measurement, use
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
1
Example:
Enable calculation:
CONF:CME:REF:P1DB ON
Query results:
FETC:CME:P1DB?
Manual operation:
See
CONFigure:CMEasurement:RESult:P3DB[:STAT]
<State>
Enables or disables the calculation of the 3
dB compression point.
To query the results of the 3
dB compression point measurement, use
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
1
Example:
Enable calculation:
CONF:CME:REF:P3DB ON
Query results:
FETC:CME:P3DB?
Manual operation:
See
CONFigure:CMEasurement:RESult:PNDB[:STAT]
<State>
Enables or disables the calculation of the n
dB compression point.
To define the compression value, use
CONFigure:CMEasurement:RESult:CVALue
To query the results of the n
dB compression point measurement, use
Parameters:
<State>
ON | OFF | 0 | 1
Configuring and Performing Measurements