![Ronde & Schwarz R&S FPL1000 Скачать руководство пользователя страница 692](http://html1.mh-extra.com/html/ronde-and-schwarz/rands-fpl1000/rands-fpl1000_user-manual_1484989692.webp)
Remote Commands
R&S
®
FPL1000
692
User Manual 1178.3370.02 ─ 09
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STOP
<Frequency>
This command defines the stop frequency of a spurious emission measurement range.
Make sure to set an appropriate span. If you set a span that is
●
smaller than the span the sweep list covers, the R&S
FPL1000 will not measure
the ranges that are outside the span - results may be invalid.
●
greater than the span the sweep list covers, the R&S
FPL1000 will adjust the start
frequency of the first range and the stop frequency of the last range to the span
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Frequency>
Numeric value.
*RST:
-2.52 MHz (range 1), 2.52 MHz (range 2), 250.0
MHz (range 3)
Default unit: Hz
Example:
LIST:RANG2:STOP 5MHZ
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:FILTer:TYPE
<FilterType>
This command selects the filter type for a spurious emission measurement range.
The EMI-specific filter types are available if the EMI (R&S
FPL1-K54) measurement
option is installed, even if EMI measurement is not active. For details see
Suffix:
<ri>
.
1..30
Selects the measurement range.
Parameters:
<FilterType>
NORMal
Gaussian filters
CFILter
channel filters
RRC
RRC filters
CISPr | PULSe
CISPR (6 dB) - requires EMI (R&S
FPL1-K54) option
Return value for query is always
PULS
.
MIL
MIL Std (6 dB) - requires EMI (R&S
FPL1-K54) option
P5
5 Pole filters
*RST:
NORM
Configuring and Performing Measurements