![Ronde & Schwarz R&S FPL1000 Скачать руководство пользователя страница 646](http://html1.mh-extra.com/html/ronde-and-schwarz/rands-fpl1000/rands-fpl1000_user-manual_1484989646.webp)
Remote Commands
R&S
®
FPL1000
646
User Manual 1178.3370.02 ─ 09
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
11.6.4.4
Selecting the Reference Channel
The following commands define the reference channel for relative ACLR measure-
ments.
[SENSe:]POWer:ACHannel:REFerence:AUTO ONCE
....................................................... 646
[SENSe:]POWer:ACHannel:REFerence:TXCHannel:AUTO
................................................ 646
[SENSe:]POWer:ACHannel:REFerence:TXCHannel:MANual
............................................. 646
[SENSe:]POWer:ACHannel:REFerence:AUTO ONCE
This command sets the channel power as the reference for relative ACLR measure-
ments.
Example:
POW:ACH:REF:AUTO ONCE
Usage:
Event
Manual operation:
See
"Setting a fixed reference for Channel Power measure-
[SENSe:]POWer:ACHannel:REFerence:TXCHannel:AUTO
<RefChannel>
This command selects the reference channel for relative measurements.
You need at least one channel for the command to work.
Parameters:
<RefChannel>
MINimum | MAXimum | LHIGhest
MINimum
Transmission channel with the lowest power
MAXimum
Transmission channel with the highest power
LHIGhest
Lowest transmission channel for lower adjacent channels and
highest transmission channel for upper adjacent channels
Example:
POW:ACH:REF:TXCH:AUTO MAX
Selects the channel with the peak power as reference channel.
Manual operation:
See
[SENSe:]POWer:ACHannel:REFerence:TXCHannel:MANual
<ChannelNumber>
This command defines a reference channel for relative ACLR measurements.
You need at least one channel for the command to work.
Configuring and Performing Measurements