General Instrument Setup
R&S
®
FPL1000
538
User Manual 1178.3370.02 ─ 09
9.8.4
How to Align the Instrument
Operating temperature
Before performing this functional test, make sure that the instrument has reached its
operating temperature (for details, refer to the data sheet).
To perform a self-alignment
Make sure no signal is connected to the RF input connector. Running a self-alignment
with a signal connected to the RF input can lead to false measurement results.
1. Press the [SETUP] key.
2. Select the "Alignment" softkey.
3. Select the "Start Self Alignment" button.
4. To abort the self-alignment process, select the "Abort Self Alignment" button.
Once the system correction values have been calculated successfully, a message
is displayed.
To display the alignment results again later
●
Press the [SETUP] key.
●
Press the "Alignment" softkey.
9.8.5
How to Align the Touchscreen
To align the touchscreen
1. Press the [Setup] key.
2. Select the "Alignment" softkey.
3. Select "Touch Screen Alignment".
A blinking cross appears in the lower left corner of the screen.
4. Touch and hold the blinking cross until it stops blinking.
Repeat this action for the crosses in the other corners.
Alignment