Remote Commands
R&S
®
FPL1000
635
User Manual 1178.3370.02 ─ 09
Example:
CONF:CME:REF MAN
CONF:CME:REF:VAL -30
Manual operation:
See
CONFigure:CMEasurement:REFerence[:MODE]
<Mode>
Determines how the reference value for the compression point measurement is
defined.
Parameters:
<Mode>
AUTO | MANual
AUTO
The reference value is determined automatically.
MANual
The reference value is defined using
*RST:
AUTO
Example:
CONF:CME:REF MAN
CONF:CME:REF:VAL -30
Manual operation:
See
CONFigure:CMEasurement:RESult:CVALue
<UserCompression>
Defines the compression value for the n
CMEasurement:RESult:PNDB[:STAT]
To query the results of the n
dB compression point measurement, use
Parameters:
<UserCompression> Distance from the linear trace to the reference trace
Default unit: DB
Example:
Enable calculation:
CONF:CME:REF:PNDB ON
Define compression value:
CONF:CME:RES:CVAL 5
Query results:
FETC:CME:PNDB?
Manual operation:
See
CONFigure:CMEasurement:RESult:P1DB[:STAT]
<State>
Enables or disables the calculation of the 1
dB compression point.
Configuring and Performing Measurements