![Ronde & Schwarz R&S FPL1000 Скачать руководство пользователя страница 509](http://html1.mh-extra.com/html/ronde-and-schwarz/rands-fpl1000/rands-fpl1000_user-manual_1484989509.webp)
General Instrument Setup
R&S
®
FPL1000
509
User Manual 1178.3370.02 ─ 09
Showing Print Colors on Display
Temporarily shows the currently selected print colors on the screen display. This func-
tion can be used as a preview for printing.
9.2.1.4
External Monitor Settings
Access
: [Setup] > "Display" > "Configure Monitor"
You can connect an external monitor (or projector) to the DVI or display port connector
on the instrument's rear panel (see the R&S
FPL1000 getting started manual).
Screen resolution and format
The touchscreen of the R&S
FPL1000 is calibrated for a 16:10 format. If you connect a
monitor or projector using a different format (e.g. 4:3), the calibration is not correct and
the screen does not react to your touch actions properly.
The touchscreen has a screen resolution of 1280x800 pixels. Most external monitors
have a higher screen resolution. If the screen resolution of the monitor is set higher
than the instrument's resolution, the application window uses an area of 1280x800 pix-
els on the monitor display. For full screen display, adjust the monitor's screen resolu-
tion.
Setup
Opens the standard Windows configuration dialog box to configure the used display
devices.
9.2.1.5
Touch(screen) Settings
Access
: [Setup] > "Display" > "Touch"
These options concern the behavior of the firmware for touch gestures on the screen.
Note that these settings remain unchanged after a channel preset.
Display Settings