General Instrument Setup
R&S
®
FPL1000
536
User Manual 1178.3370.02 ─ 09
●
If the position of the instrument has been changed and you cannot look straight on
the screen
●
If another person operates the instrument
9.8.2
Alignment Settings
Access
: [Setup] > "Alignment"
Both the instrument and the touchscreen can be aligned when necessary (see
ter 9.8.1, "Basics on Alignment"
Await Warm-Up Operation before Self Alignment
....................................................... 537
Shut down Device after Self Alignment
.......................................................................537
Starting Touch Screen Alignment
................................................................................537
Start Self Alignment
Starts recording correction data for the instrument. If the correction data acquisition
fails or if the correction values are deactivated, a corresponding message is displayed
in the status field.
For details, see
Chapter 9.8.1, "Basics on Alignment"
Note:
A running Sequencer operation is aborted when you start a self-alignment.
During self-alignment, do not connect a signal to the RF input connector. Running a
self-alignment with a signal connected to the RF input can lead to false measurement
results.
Remote command:
Alignment