![Ronde & Schwarz R&S FPL1000 Скачать руководство пользователя страница 698](http://html1.mh-extra.com/html/ronde-and-schwarz/rands-fpl1000/rands-fpl1000_user-manual_1484989698.webp)
Remote Commands
R&S
®
FPL1000
698
User Manual 1178.3370.02 ─ 09
*RST:
0
Example:
CALC:ESP:PSE:DET ON
CALC:PSE:SUBR 10
Includes up to 10 peaks per range in the list.
Manual operation:
See
CALCulate<n>:PSEarch:MARGin
<Threshold>
CALCulate<n>:PEAKsearch:MARGin
<Margin>
This command defines the threshold of the list evaluation.
Suffix:
<n>
.
Parameters:
<Margin>
Range:
-200 to 200
Default unit: dB
Example:
CALC:PSE:MARG 100
Sets the threshold to 100 dB.
Manual operation:
See
CALCulate<n>:PSEarch:PSHow
<State>
CALCulate<n>:PEAKsearch:PSHow
<State>
This command turns the peak labels in the diagram on and off.
Peak labels are blue squares.
Suffix:
<n>
.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:PSE:PSH ON
Marks all peaks with blue squares.
Manual operation:
See
CALCulate<n>:PSEarch:SUBRanges
<NumberPeaks>
CALCulate<n>:PEAKsearch:SUBRanges
<NumberPeaks>
This command defines the number of peaks included in the peak list.
After this number of peaks has been found, the R&S
FPL1000 stops the peak search
and continues the search in the next measurement range.
Suffix:
<n>
.
Configuring and Performing Measurements