![Ronde & Schwarz R&S FPL1000 Скачать руководство пользователя страница 146](http://html1.mh-extra.com/html/ronde-and-schwarz/rands-fpl1000/rands-fpl1000_user-manual_1484989146.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
146
User Manual 1178.3370.02 ─ 09
The R&S
FPL1000 sets the optimum RF attenuation and the reference level for the
power in the transmission channel to obtain the maximum dynamic range. The fol-
lowing figure shows the result of the measurement.
Figure 7-23: Measuring the relative adjacent channel power on a W-CDMA uplink signal
The R&S
FPL1000 measures the power of the individual channels. A root raised
cosine filter with the parameters
α
=
0.22 and chip rate 3.84
Mcps (=
receive filter
for W-CDMA) is used as channel filter.
Measurement Example 3 – Measuring the Intrinsic Noise of the R&S
FPL1000
with the Channel Power Function
Noise in any bandwidth can be measured with the channel power measurement func-
tions. Thus the noise power in a communication channel can be determined, for exam-
ple.
If the noise spectrum within the channel bandwidth is flat, the noise marker can be
used to determine the noise power in the channel by considering the channel band-
width. However, in the following cases, the channel power measurement method must
be used to obtain correct measurement results:
●
If phase noise and noise that normally increases towards the carrier is dominant in
the channel to be measured
●
If there are discrete spurious signals in the channel
Test setup:
► Leave the RF input of the R&S
FPL1000 open-circuited or terminate it with 50 Ω.
Procedure:
1. Preset the R&S
FPL1000.
Measurements and Results