Remote control commands
R&S
®
NRP18S-xx
67
User Manual 1178.3686.02 ─ 07
Trigger level
Power Trigger event
Timeslot 1
Timeslot 2
Timeslot 3
Timeslot width
Time
Start of fence
Length of fence
Figure 8-3: Timeslot parameters
Triggering a timeslot measurement
In timeslot mode, internal and external trigger events from the signal are evaluated
depending on the settings of the
parameter. It is essential to define
parameter to ensure that the beginning of the first slot to be mea-
sured coincides with the delayed trigger point.
Defining a time interval for the measurement
At the beginning and at the end of the measurement interval, you can define time inter-
vals that are excluded from the measurement, see
Chapter 8.8.3, "Excluding intervals"
Remote commands:
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:COUNt
................................................................ 67
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:WIDTh
.................................................................68
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID:OFFSet[:TIME]
..............................68
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID:TIME
............................................68
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID[:STATe]
........................................ 68
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:COUNt
<count>
Sets the number of simultaneously measured timeslots in timeslot mode. See
Parameters:
<count>
Range:
1 to 128
*RST:
8
Configuring the measurement modes