Parameters:
<hysteresis>
Range:
0.00 to 10.00
*RST:
0.00
Default unit: DB
TRIGger:IMMediate
Causes a generic trigger event. The power sensor leaves the waiting for trigger state
immediately, irrespective of the trigger source and the trigger delay, and starts the
measurement.
This command is the only way to start a measurement if the trigger source is set to
hold,
HOLD
. Only one measurement cycle is executed, irrespective
of the averaging factor.
Usage:
Event
TRIGger:LEVel
<level>
Effective only if
INTernal
.
Sets the trigger threshold for internal triggering derived from the test signal.
If an S-parameter device and/or the offset correction are active, the currently effective
trigger threshold and a trigger threshold to be input are referenced to the appropriately
shifted sensor data.
If the S-parameter device and/or the offset correction are disabled, the trigger thresh-
old and its input limits are adjusted as necessary.
If you enter a value without unit, the unit is defined by
Parameters:
<level>
Range:
1.0e-7 to 200.0e-3
*RST:
1.0e-6
Default unit: Watts
TRIGger:LEVel:UNIT
<unit>
Sets the unit of the trigger level if this value is entered without a unit. See also
Parameters:
<unit>
DBM | W | DBUV
*RST:
W
TRIGger:SENDer:PORT
<sender_port>
Selects the port where the power sensor outputs its own trigger event in case it is trig-
ger sender. See
for more information.