![Ronde & Schwarz 1424.6721K02 Скачать руководство пользователя страница 45](http://html1.mh-extra.com/html/ronde-and-schwarz/1424-6721k02/1424-6721k02_user-manual_1484980045.webp)
Remote control commands
R&S
®
NRP18S-xx
45
User Manual 1178.3686.02 ─ 07
8.5.1
Starting and ending a measurement
ABORt
Immediately interrupts the current measurement. If the measurement has been started
as a single measurement (
), the power sensor goes into the
idle state. However, if a continuous measurement is in progress (
ON
), the trigger system of the power sensor enters the waiting for trigger
state. When the trigger condition is met, a new measurement is immediately started.
See also
Chapter 8.5, "Controlling the measurement"
Usage:
Event
INITiate:ALL
INITiate[:IMMediate]
Starts a single measurement cycle. The power sensor changes from the idle state to
the waiting for trigger state. When the trigger condition is fulfilled, the power sensor
begins the measurement. Depending on the number of trigger events that are required,
e.g. for averaging, the power sensor enters the waiting for trigger state several times.
Once the entire measurement is completed, a measurement result is available, and the
power sensor enters the idle state again.
Use this command only after the continuous measurement mode has been disabled
using
OFF
.
See also
Chapter 8.5, "Controlling the measurement"
Example:
See
Chapter 9.3, "Performing a buffered continuous average
Usage:
Event
INITiate:CONTinuous
<state>
Enables or disables the continuous measurement mode. In continuous measurement
mode, the power sensor does not go into the idle state after a measurement has been
completed, but immediately executes another measurement cycle.
See also
Parameters:
<state>
ON
Measurements are performed continuously. If a measurement is
completed, the sensor does not return to the idle state but enters
the waiting for trigger state again.