Remote control commands
R&S
®
NRP18S-xx
104
User Manual 1178.3686.02 ─ 07
SYSTem:PARameters:DELTa?
Returns an XML-output containing all commands that differ from the defined default
status set by
The commands are accompanied by the same information as for
.
Usage:
Query only
SYSTem:PRESet
Resets the power sensor.
The command essentially corresponds to the
*RST
command, with the exception that
the settings of the following commands are persistently held:
INITiate:CONTinuous
SENSe:AVERage:TCONtrol
SENSe:TRACe:AVERage:TCONtrol
Usage:
Event
SYSTem:REBoot
Reboots the power sensor.
Usage:
Event
SYSTem:RESTart
Restarts the firmware of the power sensor.
Usage:
Event
SYSTem:RUTime
<update_time>
Effective only in the NRP legacy mode. Relevant only in continuous measurement
mode,
Sets the result update time. That is the maximum rate in which the power sensor can
output measurement results.
Parameters:
<update_time>
Range:
0.0 to 10.0
*RST:
0.1
Default unit: Seconds
SYSTem:SERRor?
Queries the next static error, if available.
Configuring the system