Remote control commands
R&S
®
NRP18S-xx
66
User Manual 1178.3686.02 ─ 07
Defining a time interval for the measurement
At the beginning and at the end of the measurement interval, you can define time inter-
vals that are excluded from the measurement, see
Chapter 8.8.3, "Excluding intervals"
Remote commands:
[SENSe<Sensor>:][POWer:]BURSt:DTOLerance
................................................................ 66
[SENSe<Sensor>:][POWer:]BURSt:LENGth?
..................................................................... 66
[SENSe<Sensor>:][POWer:]BURSt:DTOLerance
<tolerance>
Sets the drop-out tolerance, a time interval in which the pulse end is only recognized if
the signal level no longer exceeds the trigger level. See
.
Parameters:
<tolerance>
Range:
0.00 to 0.30
*RST:
1.000e-6
Default unit: Seconds
[SENSe<Sensor>:][POWer:]BURSt:LENGth?
Queries the length of a burst (pulse interval), the time between the trigger point of the
measurement and the time the trigger logic detects the end of the pulse. See
Usage:
Query only
8.7.3
Timeslot measurement
The timeslot mode is used to measure the average power of a definable number of
successive timeslots within a frame structure with equal spacing. The measurement
result is an array with the same number of elements as timeslots. Each element repre-
sents the average power in a particular timeslot.
Configuring the measurement modes