Remote control commands
R&S
®
NRP18S-xx
60
User Manual 1178.3686.02 ─ 07
8.6.2
Setting the result format
The
FORMat
subsystem sets the format of numeric data (measured values) that is
exchanged between the remote control computer and the power sensors if high-level
measurement commands are used.
Remote commands:
FORMat:BORDer
<border>
Selects the order of bytes in 32-bit or 64-bit binary data.
Parameters:
<border>
NORMal | SWAPped
NORMal
The 1st byte is the least significant byte, the 4th/8th byte the
most significant byte.
Fulfills the Little Endian (little end comes first) convention, used
by x86/x64 CPUs, for example.
SWAPped
The 1st byte is the most significant byte, the 4th/8th byte the
least significant byte.
Fulfills the Big Endian (big end comes first) convention.
*RST:
NORMal
Example:
FORM:BORD NORM
FORMat:SREGister
<sregister>
Specifies which format is used for the return value of
Parameters:
<sregister>
ASCii | HEXadecimal | OCTal | BINary
*RST:
ASCii
Example:
FORM:SREG ASC
FORMat[:DATA]
[<data,length>, <length>]
Specifies how the R&S NRP18S-xx sends the numeric data to the controlling host/
computer.
Parameters:
<data,length>
<REAL,32 | 64>
REAL
Block of binary values, 32-bit or 64-bit each; so-called "SCPI
definite length block"
Configuring results