![Ronde & Schwarz 1424.6721K02 Скачать руководство пользователя страница 132](http://html1.mh-extra.com/html/ronde-and-schwarz/1424-6721k02/1424-6721k02_user-manual_1484980132.webp)
Remote control basics
R&S
®
NRP18S-xx
132
User Manual 1178.3686.02 ─ 07
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
0
Sensor initializing
0
0
0
0
0
0
0
0
0
0
0
0
0
0
+
Figure 10-11: Operation sense status register
Querying the register:
●
STATus:OPERation:SENSe:CONDition?
●
STATus:OPERation:SENSe[:SUMMary][:EVENt]?
Table 10-13: Used operation sense status bits and their meaning
Bit no.
Short description
Bit is set if
1
Sensor initializing
Sensor is being initialized.
10.2.5.5
Operation lower limit fail status register
The
CONDition
registers contain information whether a measured value is below a
configured lower limit. The
EVENt
registers contain information whether a measured
value dropped below a limit value since the last query.