![Ronde & Schwarz 1424.6721K02 Скачать руководство пользователя страница 58](http://html1.mh-extra.com/html/ronde-and-schwarz/1424-6721k02/1424-6721k02_user-manual_1484980058.webp)
Remote control commands
R&S
®
NRP18S-xx
58
User Manual 1178.3686.02 ─ 07
If the power sensor is the trigger sender, it can output its trigger event either on the
EXTernal<1>
or
EXTernal2
.
If the power sensor triggers itself, the trigger source of the power sensor must be
assigned to the other external port, as shown in the examples.
Parameters:
<sender_port>
EXT1 | EXTernal1 | EXT2 | EXTernal2
*RST:
EXT1
Example:
TRIG:SEND:PORT EXT1
TRIG:SOUR EXT2
TRIG:SEND:STAT ON
Example:
TRIG:SEND:PORT EXT2
TRIG:SOUR EXT1
TRIG:SEND:STAT ON
TRIGger:SENDer:STATe
<state>
Enables or disables the trigger sender mode of the power sensor. In this state, the
power sensor can output a digital trigger signal in sync with its own trigger event.
If enabled, select the output port for the trigger signal using
.
Typically, the trigger sender uses its internal trigger source. But you can also trigger the
trigger sender externally, because the power sensor has got two external trigger con-
nectors. If you trigger the trigger sender externally, use
EXTernal1
as external trigger
input port (trigger source) and
EXTernal2
as trigger sender output port or vice versa.
Parameters:
<state>
*RST:
OFF
TRIGger:SLOPe
<slope>
Effective only if
INTernal
or
EXTernal
.
Determines which edge of the envelope power, with internal triggering, or increasing
voltage, with external triggering, is used for triggering.
Parameters:
<slope>
POSitive | NEGative
POSitive
Rising edge
NEGative
Falling edge
*RST:
POSitive
TRIGger:SOURce
<source>
Selects the source for the trigger event detector.