Remote control commands
R&S
®
NRP18S-xx
106
User Manual 1178.3686.02 ─ 07
Parameters:
<update_time>
Range:
0.0 to 10.0
*RST:
10e-3
Default unit: Seconds
SYSTem:TLEVels?
Queries the possible power test levels of the power sensor.
Usage:
Query only
SYSTem:TRANsaction:BEGin
Starts a series of settings.
Usage:
Event
SYSTem:TRANsaction:END
Ends a series of settings.
Usage:
Event
SYSTem[:SENSor]:NAME
<sensorname>
Sets the name of the power sensor according to your requirements. The specified
name is displayed in the web user interface of the network power sensors.
The name that you specify here is independent from the hostname of the power sen-
sor. However, if you do not specify a name, the hostname is used as default.
Parameters:
<sensorname>
Example:
SYST:NAME "InputModule-X5"
SYSTem:VERSion?
Queries the SCPI version that the command set of the power sensor complies with.
Example:
SYST:VERS?
Query
1999.0
Response: SCPI version from 1999.
Usage:
Query only
8.13
Using the status register
Further information:
Using the status register