Remote control commands
R&S
®
NRP18S-xx
68
User Manual 1178.3686.02 ─ 07
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:WIDTh
<width>
Sets the length of the timeslot. See
.
Parameters:
<width>
Range:
10.0e-6 to 0.10
*RST:
1.000e-3
Default unit: Seconds
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID:OFFSet[:TIME]
<time>
Determines the distance from the start of the timeslots to the start of the interval to be
blanked out. See
.
Parameters:
<time>
Range:
0.00 to 0.10
*RST:
0.00
Default unit: Seconds
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID:TIME
<time>
Sets the length of the time interval in the timeslots to be excluded from the measure-
ment. See
. The parameter applies to each individual timeslot.
Note:
Even if the exclusion interval exceeds the timeslot because, for example, its right
limit is outside the timeslot, correct results are obtained. In the extreme case, where
the interval length has been set to a value greater than the timeslot length, 0
W is out-
put as the measured power. No error message is output.
Parameters:
<time>
Range:
0.00 to 0.10
*RST:
0.00
Default unit: Seconds
[SENSe<Sensor>:][POWer:]TSLot[:AVG][:EXCLude]:MID[:STATe]
<state>
Enables or disables the blanking out of time intervals in the timeslots.
Parameters:
<state>
*RST:
OFF
8.7.4
Trace measurement
The trace measurement determines the course of power over a defined time. During
the measurement time set by
measurements are performed. The result is returned as an array of values with a size
predefined by
. The length of an individual mea-
surement(-point) is determined from the ratio of measurement time and measurement
points. The entire result is called a "trace". Each trace must be triggered separately.
Configuring the measurement modes