Remote control commands
R&S
®
NRP18S-xx
55
User Manual 1178.3686.02 ─ 07
TRIGger:ATRigger[:STATe]
<state>
Controls the automatic trigger function. If enabled, an artificial trigger is generated if the
delay time has elapsed after the measurement start and no trigger event has occurred.
The delay time is set using
Parameters:
<state>
*RST:
OFF
TRIGger:COUNt
<count>
Sets the number of measurement cycles to be performed when the measurement is
started using
.
This number equals the number of results that can be obtained from the sensor after a
single measurement. As long as the defined number of measurements is not executed,
the sensor automatically initiates another measurement internally when the current
result is available.
This command is particularly useful in conjunction with buffered measurements. For
example, to fill a buffer with a predefined size with measurements that have been trig-
gered externally or by
without having to start the measurement multiple times.
Parameters:
<count>
Range:
1 to 8192
*RST:
1
Example:
See
Chapter 9.3, "Performing a buffered continuous average
TRIGger:DELay
<delay>
Sets the delay between the trigger event and the beginning of the actual measurement
(integration).
Parameters:
<delay>
Range:
-5.0 to 10.0
*RST:
0.0
Default unit: s
TRIGger:DELay:AUTO
<state>
If
ON
is set, no measurement is started until the power sensor
has settled. For this purpose, the delay value is automatically determined.
If a longer period is set using
, the automatically determined delay is
ignored.
Parameters:
<state>
*RST:
OFF
Controlling the measurement