Example:
TRAC:DATA?
Usage:
Query only
[SENSe<Sensor>:]TRACe:MPWidth?
Queries the attainable time resolution in trace mode. The result is the smallest possible
distance between two pixels, i.e. it is the smallest time interval that can be assigned to
a pixel.
Usage:
Query only
[SENSe<Sensor>:]TRACe:OFFSet:TIME
<time>
Adds an offset to the beginning of the trace sequence. Thus, the trace in the result dis-
play is moved in positive or negative x-direction. If you measure with more than one
sensor, you can use this offset to arrange the traces to each other. The start of record-
ing relative to the trigger event is set using
.
Parameters:
<time>
Range:
- (500 ms + TRGger:DELay) to 100 s
*RST:
0.0
Default unit: Seconds
Example:
TRAC:OFFS:TIME 1.0
[SENSe<Sensor>:]TRACe:POINts
<points>
Sets the number of required values per trace sequence.
Parameters:
<points>
Range:
1 to 100000
*RST:
260
[SENSe<Sensor>:]TRACe:REALtime
<state>
Available in trace mode.
If disabled, each measurement from the power sensor is averaged. If enabled, only
one sampling sequence per measurement is recorded, thus increasing the measure-
ment speed. With a higher measurement speed, the measured values of an individual
measurement are immediately delivered.
The averaging filter is not used, so the following settings are ignored:
●
[SENSe<Sensor>:]TRACe:AVERage[:STATe]
●
[SENSe<Sensor>:]TRACe:AVERage:COUNt
Parameters:
<state>
*RST:
OFF