![Ronde & Schwarz 1424.6721K02 Скачать руководство пользователя страница 129](http://html1.mh-extra.com/html/ronde-and-schwarz/1424-6721k02/1424-6721k02_user-manual_1484980129.webp)
Remote control basics
R&S
®
NRP18S-xx
129
User Manual 1178.3686.02 ─ 07
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
0
Sensor calibrating
0
0
0
0
0
0
0
0
0
0
0
0
0
0
+
Figure 10-8: Operation calibrating status register
Querying the register:
●
STATus:OPERation:CALibrating:CONDition?
●
STATus:OPERation:CALibrating[:SUMMary][:EVENt]?
Table 10-10: Used operation calibrating status bits and their meaning
Bit no.
Short description
Bit is set if
1
Sensor calibrating
Sensor is being calibrated.
10.2.5.2
Operation measuring status register
The
CONDition
register contains information whether a power sensor is measuring.
The
EVENt
register contains information whether a measurement was started or com-
pleted since the last query.