Remote control commands
R&S
®
NRP18S-xx
63
User Manual 1178.3686.02 ─ 07
[SENSe<Sensor>:][POWer:][AVG:]APERture
......................................................................63
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:CLEar
...............................................................63
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:COUNt?
............................................................63
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:DATA?
.............................................................. 63
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE
.................................................................64
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:STATe
...............................................................64
[SENSe<Sensor>:][POWer:][AVG:]FAST
............................................................................ 64
[SENSe<Sensor>:][POWer:][AVG:]SMOothing:STATe
.......................................................... 65
[SENSe<Sensor>:][POWer:][AVG:]APERture
<integration_time>
Sets the duration of the sampling window in continuous average mode. During this
time interval, the average signal power is measured.
The minimum value is implemented for fast unchopped continuous average measure-
ments. See also
Chapter 9.2.2, "Triggered fast unchopped continuous average mea-
Parameters:
<integration_time>
Range:
8.0e-6 to 2.00
*RST:
0.02
Default unit: Seconds
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:CLEar
Clears the contents of the result buffer in continuous average mode.
Example:
BUFF:CLE
Usage:
Event
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:COUNt?
Available in continuous average mode.
Queries the number of results that are currently stored in the result buffer.
Example:
BUFF:COUN?
Usage:
Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:DATA?
Queries the results of the continuous average result buffer and returns them even if the
buffer is not full.
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
the buffer is full.
Usage:
Query only
Configuring the measurement modes