Remote control commands
R&S
®
NRP18S-xx
77
User Manual 1178.3686.02 ─ 07
8.8.2
Setting the frequency
The frequency of the signal to be measured is not automatically determined. For ach-
ieving better accuracy, the carrier frequency of the applied signal must be set.
[SENSe<Sensor>:]FREQuency
<frequency>
Transfers the carrier frequency of the RF signal to be measured. This frequency is
used for the frequency-response correction of the measurement result.
The center frequency is set for broadband signals, e.g. spread-spectrum signals, multi-
carrier signals.
Parameters:
<frequency>
Range:
0.0 to 110.0e9
*RST:
50.0e6
Default unit: Frequency
Example:
FREQ 10000
8.8.3
Excluding intervals
In the burst average and timeslot modes, you can define a time interval at the begin-
ning or at the end of an integration interval that is excluded from the measurement
result. See in
.
p
t
2
1
3
4
Figure 8-6: Excluding intervals
1 = Trigger level
2 = Excluding interval at the beginning
3 = Excluding interval at the end
4 = Integration interval
Remote commands:
[SENSe<Sensor>:]TIMing:EXCLude:STARt
........................................................................ 78
[SENSe<Sensor>:]TIMing:EXCLude:STOP
.........................................................................78
Configuring basic measurement parameters