Remote Control Commands
R&S
®
NRPxxA(N)
121
User Manual 1177.6017.02 ─ 05
For more information, see
.
Parameters:
<mode>
USER | SENSor
*RST:
SENSor
SYSTem:MINPower?
Queries the lower power measurement limit.
[SENSe<Sensor>:]CORRection:SPDevice:STATe
is set to
ON
. The lower measurement limit refers to the sensor or to the combination of a sensor
and the components connected ahead of it.
Use this query to determine a useful resolution for the result display near the lower
measurement limit.
Usage:
Query only
SYSTem:PARameters?
Lists all commands with default values, limits and ranges.
Usage:
Query only
SYSTem:PARameters:DELTa?
Lists all commands that differ from the defined default status set by
The commands are output with default values, limits and ranges.
Usage:
Query only
SYSTem:PRESet
Triggers a sensor reset.
The command essentially corresponds to the
*RST
command, with the exception that
the settings of the following commands are persistently held:
INITiate:CONTinuous
SENSe:AVERage:TCONtrol
Usage:
Event
SYSTem:REBoot
Reboots the sensor.
Usage:
Event